I. General Preparation and Stability
Confirm the status of the instrument and the integrity of the accessories: Check the power cord, test leads, test clips/probes, and grounding to ensure they are reliable and undamaged, and there is no breakage or loosening.
Read the manual: Familiarize yourself with the definitions of each port, the rated values, the test modes, and the usage methods of the accompanying fixtures (such as four-terminal fixtures, scanning fixtures).
Power-off wiring: After the equipment under test is powered off and the capacitor energy storage components have been fully discharged, the wiring and disconnection operations can be carried out.
Reliable grounding: Ensure a secure connection between the instrument's protective grounding terminal and the protective ground of the tested equipment to prevent risks to personnel and equipment.
II. Standard wiring steps according to test type
Ratio/Group/Phase Angle Test (Ratio Tester)
Single-phase: The yellow/green terminals on the high-voltage side are connected to the high-voltage terminals A and N respectively; the yellow/green terminals on the low-voltage side are connected to the low-voltage terminals a and n.
Three-phase: The yellow/green/red colors on the high-voltage side correspond to A/B/C; the yellow/green/red colors on the low-voltage side correspond to a/b/c. The phases must be in the same sequence for the test to be conducted.
Special Connection:
Z-type: The high-voltage side has four wires connected to A, B, C, and N; the low-voltage side has three wires connected to a, b, and c.
Scott: The high-voltage side is connected to A, B, and C; the low-voltage side is connected to α, β, and n.
Inverse Scott: Connect α, β, n to the high-voltage side; connect a, b, c to the low-voltage side.
After the connection is completed, power on and enter the test menu to select the corresponding mode to start the measurement.
Comprehensive parameter testing (including LCR/turn ratio/inductance leakage/direct current resistance, etc., with test fixture)
Power supply and fixture: Connect the instrument power supply to 220 VAC and turn on the power; Connect the SCANNER cable of the test fixture to the SCANNER interface on the rear panel of the instrument.
Four-terminal connection: Firmly connect the UNKNOWN four-terminal cable of the test end to the interface of the test stand to ensure reliable contact during low-resistance measurement.
Clamping of the test piece: According to the silk-screening on the fixture, reliably press-fit or securely clamp each pin of the transformer to avoid shaking and contact resistance issues.
Capacity and loss characteristic test (capacity / active load / no-load / short-circuit test)
Active capacity/load loss: For the high-voltage side, use three test clips (yellow/green/red), each clip having two wires connected to the corresponding colored current terminals Ia, Ib, Ic (thick wires) and voltage terminals Ua, Ub, Uc (thin wires), with the clip heads clamped onto the respective connection terminals on the high-voltage side; for the low-voltage side, use a dedicated shorting wire to make a good short connection.
Passive test (no-load / short-circuit): Connect the power supply of the external three-phase voltage regulator to the characteristic test terminals on the instrument panel, and follow the wiring and settings for the test items.
Key points for parameter setting: Set the current temperature, rated voltage on the high-voltage side, transformer type, tap position, etc., and then proceed to enter the corresponding test items to start.
III. Rapid Self-Check After Wiring Completion
Contact reliability: Pull the test wire gently to confirm no loosening; during the four-terminal test, ensure that the current circuit and the voltage circuit are independent of each other and the contact is stable.
Parameter presetting: Input the nameplate information (such as rated ratio, rated voltage, total number of taps/spacing, transformer type) in the instrument. If necessary, select the Y/D/Z connection method.
Testing and Storage: First, conduct short-term point measurements to observe the stability of the data, then proceed with the formal test; if necessary, save or print the results for future reference.
IV. Common Errors and Troubleshooting
Low-voltage side not short-circuited (capacity/load test): This will cause measurement distortion. It is essential to use a dedicated short-circuiting wire to reliably short-circuit the low-voltage side.
Poor contact/oxidation: Causes excessive DC resistance and poor repeatability. Clean and tighten the connection terminals.
Misuse of test mode: For capacity/loss test, use the no-load/short-circuit port or the unconnected voltage regulator. Follow the instructions to switch to the correct mode and terminals.