This is the fundamental starting point for selecting equipment. Different test samples have completely different requirements for DC high-voltage generators.
Test sample type Primary test items Key requirements for the equipment
Power cables (such as: cross-linked polyethylene XLPE, oil-paper insulation) Direct current withstand test, leakage current test 1. High voltage level (usually 2.5-3 times U₀)
2. Large capacity (current) (usually requires several mA to 10 mA or more to overcome capacitive current and detect leakage)
3. Good stability, small ripple coefficient
Zinc oxide arresters (MOA) Direct current 1mA reference voltage (U₁mA) and leakage current at 0.75U₁mA 1. Moderate voltage level (select according to the rated voltage of the arrester)
2. High accuracy requirements, especially the current measurement accuracy must reach the micro-ampere level (µA)
3. Can be smoothly adjusted from 0-100% rated voltage
Generators, motors Direct current withstand test, leakage current test 1. High voltage level
2. Extremely large capacity (due to the large geometric capacitance of the motor, the charging current is large, usually requiring several mA or even hundreds of mA)
3. Good overcurrent protection function
Transformers, bushings, transformers Direct current withstand test, leakage current test The voltage and capacity are determined according to the equipment's voltage level, and usually require a stable high-voltage source.
Once the test object is identified, one can focus on the technical specifications of the DC high-voltage generator.
Rated Voltage (kV):
Based on the test standards of the tested item (such as DL/T 596), determine the required maximum test voltage.
Selection principle: The rated voltage of the equipment should be slightly higher than the maximum test voltage you require, leaving a certain margin (e.g. 10-20%), but it should not be too large as otherwise the equipment would be bulky and expensive.
Rated current (mA) and power (W):
This is the core indicator for measuring the load-bearing capacity of the equipment. Power = Voltage × Current.
How to calculate the required current
For capacitive loads (such as cables): The initial charging current is very high. The required current is greater than 2πfCU × 1.2 (where C is the capacitance of the tested item, which can be obtained from the manufacturer's data or measured by an ohmmeter).
For resistive loads (such as lightning arresters): The main consideration is leakage current. Typically, a model with 2mA or 3mA leakage current can meet the requirements of most MOA tests.
Selection principle: Prefer a larger one over a smaller one. If the current is set too low, when boosting the voltage for large-capacitor equipment, the voltage may fail to rise or be pulled down. For cable and generator tests, it is recommended to choose a model with 5mA or higher.
Voltage and current measurement accuracy:
The higher the accuracy, the more reliable the reading. The voltage and current measurement accuracy of mainstream high-quality equipment is usually within 1.0% to 1.5%. For precise research or arbitration tests, higher accuracy (such as 0.5%) is required.
Ripple coefficient:
It refers to the proportion of the AC component in the output voltage. The smaller the ripple coefficient, the purer the DC voltage, and the more accurate the test results will be.
The national standard stipulates that the ripple coefficient of the DC high-voltage generator used for pressure resistance and leakage tests should not exceed 3%. High-quality equipment can achieve a ripple coefficient of 1% or even lower.
Protection function:
Overvoltage protection: It can be set. When the voltage exceeds the set value, it will automatically cut off.
Overcurrent protection: It can be set. When the leakage current exceeds the set value, it will automatically cut off.
Zero position protection: Prevents the generation of high voltage at the moment of power-on.
Grounding protection: Ensure that the equipment cannot be started if it is not properly grounded.
These protective functions are essential for ensuring the safety of equipment and personnel, and none of them can be omitted.
Brand and Quality:
Choose brands that have a good reputation and recognition within the power and electrical industries. For example, Wuhan Hezhong Electric. The products of these brands are more reliable in terms of stability, reliability and safety.
Check whether the necessary type test reports and verification certificates are available.
After-sales Service:
Verify whether the supplier can provide timely technical support and maintenance services.
Understand the warranty period and calibration services of the equipment.
Budget:
Within the budget limit, priority should be given to meeting the core requirements such as voltage, current, accuracy and protection functions.
Fully automatic and microcomputer-controlled equipment is more expensive than manual equipment, but the increased efficiency and reliability it offers are well worth it.